Description
Medin CNO® CPAP Driver – Advanced Respiratory Support for Neonates
The medinCNO® CPAP Driver is a precision-engineered device designed for administering Continuous Positive Airway Pressure (CPAP) therapy to premature infants and newborns. Paired with the Medijet® nCPAP generator, this device is critical in neonatal respiratory care and must be used under the supervision of trained clinical staff.
Key Features & Benefits Medin CNO®
✅ High-Precision CPAP Therapy
-
Measurement Range: 0 to 18 mbar, accurate within ±1.3 mbar
-
Leakage Compensation: Leak-Assist function for stable CPAP pressure
-
Automatic Apnea Detection: With push flow support for accurate respiration regulation
✅ Optimized for Newborns
-
Designed for neonatal care, providing consistent respiratory support
-
Adjustable Oxygen Concentration (FiO2), 21% – 100%
-
Oscillation Frequency of 5 to 20 Hz for precise pressure control
✅ Innovative Data Monitoring
-
7-inch color display with real-time data on CPAP, flow, FiO2, RR, push frequency
-
Redundant sensor technology for double assurance of patient data accuracy
-
USB & RS232 ports for data export and live trend monitoring
✅ Reliable & Safe
-
Dual power supply with internal battery (up to 3 hours of runtime) and external power support
-
Mechanical overpressure valve for safety
-
Electronic shut-off valve to stop flow supply in case of errors
Technical Specifications medin CNO®
-
Dimensions: 29 cm x 23.5 cm x 18.5 cm
-
Weight: 4.75 kg
-
Power Supply: 14.4 V DC (internal battery), rechargeable
-
External Power Supply: 100-240 V AC, 50-60 Hz
-
Flow Range: 0 l/min to 15 l/min, working range: 4 l/min to 10 l/min
-
Accuracy: ±1 l/min (in working range), ±2 l/min (outside working range)
-
CPAP Pressure Measurement: 0 to 18 mbar
-
Oxygen Concentration (FiO2): 21 to 100%, accuracy ±3%
-
Alarms: Visual (LED) & acoustic alarms, adjustable settings
-
Operating Time: Up to 4 weeks of continuous operation without restarting
Modes of Operation medin CNO®
-
CPAP Mode: Standard CPAP with Leak-Assist
-
Apnea CPAP Mode: Automatically generates pushes during apnea detection
-
NIPPV Mode: Bi-level CPAP with I:E ratio measurement
-
Oscillation Mode: Superimposed high-frequency oscillation for pressure support
-
SNIPPV Mode: Synchronizes with the patient’s inspiration for precise support
Regulatory Compliance & Certification
-
EC Classification: Class IIb medical device
-
IEC Certifications: 60601-1, 60601-1-2, 60601-1-8
-
IP Protection Class: IP20
-
Certified Quality: EN ISO 13485, EN ISO 9001